4 GMC-I Messtechnik GmbH
3 Performing Measurements
3.1 Switching the Instrument On and Off, Energy-Saving
Mode, Automatic Shutdown
The instrument is switched on by pressing and holding the START
key. Briefly press the START key twice in order to switch the instru-
ment off, during which no voltage may be applied to the test
probes! The instrument is switched to the standby mode after
several seconds (reduced brightness), if none of the keys has
been pressed and no voltage is applied to the test probes. The
instrument is switched out of the standby mode (i.e. back to full
brightness) by pressing any key or applying voltage to the test
probes. The instrument is shutdown automatically if it has
remained inactive for about 1 minute, i.e. if no keys have been
pressed and no voltage has been applied to the test probes dur-
ing this time.
3.2 Instructions and Principles with Validity for
All Measurements
The desired functions or parameters are selected with the
R
ISO/USPD, FUNC, UTEST and UTEST keys. Measurement is
triggered by pressing the START key. All selected functions or
parameters remain valid until they are changed.
If (interference) voltage of greater than 10 V is applied to the
test probes, this is indicated at the display in the UIN field
along with the measured value, as well as the “!” symbol. If
this is the case, the START key is disabled.
Bild 3.1 (Interference) Voltage Display
If battery voltage is too low (only the red field is lit up in the
battery symbol), measurement cannot be started. After press-
ing the START key, the depleted battery symbol appears for
about 1 second (see figure below). Replace the batteries as
described in section 5.1 on page 7.
When measuring high insulation resistance, the measurement
cables should be laid unobstructed in the room, or placed
upon a material with good insulating properties.
Contact the device under test with the test probe before trig-
gering measurement with the START key, thus causing any
possible interference voltage to appear at the display.
Do not remove the test probes prematurely during the course
of the measurement, because this may lead to distorted mea-
surement results.
After insulation measurement has been completed, any remaining
residual voltage is displayed as U
IN , which may result from cable
capacitance. Contact to the device under test must be main-
tained for as long as a capacitive DUT is being discharged via the
test instrument’s internal resistor. The falling voltage value can be
observed directly at the U
IN display. Do not disconnect the DUT
until voltage U
IN is less than 25 V!
3.3 Measuring Functions
Attention!
Each time before measuring insulation resistance or surge
protection, make sure that the device under test is voltage-
free!
Do not touch the device under test or the test probes during
measurement, and for a short time thereafter as long the DUT
is being discharged.
The DUT may be charged with as much as 1000 V during
measurement, for which reason contact to the DUT must be
maintained after measurement has been completed until it is
fully discharged! Residual voltage appears at the display along
with the “!” warning symbol.
Do not remove the test probes until voltage has dropped to a
safe level and the “!” warning symbol has been cleared from
the display.
All consuming devices must be disconnected and the
switches must be turned on when measuring insulation resis-
tance between conductors.
3.3.1 Voltage Measurement
Connect the test instrument to the device under test.
Figure 3.4: Connection Example for Voltage Measurement
If a voltage of greater than about 10 V is present in the mea-
sured electrical circuit, its value is displayed in the U
IN field and
the “~” symbol appears for AC or polarity is displayed for DC.
The “+” symbol lights up if the positive pole is connected to
the fixed test probe, and “–” appears if the negative pole is
connected to the fixed test probe. “!” appears at the same
time as a warning. The START key is disabled!
Figure 3.2: Low Battery
Voltage
Figure 3.3: Low Battery
Voltage: Display
After Pressing the
START Key
Figure 3.5: Display of (Inter-
ference) Voltage
(Riso function)
Figure 3.6: Voltage
Measurement
(U
SPD function)
!
4

Hulp nodig? Stel uw vraag in het forum

Spelregels

Misbruik melden

Gebruikershandleiding.com neemt misbruik van zijn services uitermate serieus. U kunt hieronder aangeven waarom deze vraag ongepast is. Wij controleren de vraag en zonodig wordt deze verwijderd.

Product:

Bijvoorbeeld antisemitische inhoud, racistische inhoud, of materiaal dat gewelddadige fysieke handelingen tot gevolg kan hebben.

Bijvoorbeeld een creditcardnummer, een persoonlijk identificatienummer, of een geheim adres. E-mailadressen en volledige namen worden niet als privégegevens beschouwd.

Spelregels forum

Om tot zinvolle vragen te komen hanteren wij de volgende spelregels:

Belangrijk! Als er een antwoord wordt gegeven op uw vraag, dan is het voor de gever van het antwoord nuttig om te weten als u er wel (of niet) mee geholpen bent! Wij vragen u dus ook te reageren op een antwoord.

Belangrijk! Antwoorden worden ook per e-mail naar abonnees gestuurd. Laat uw emailadres achter op deze site, zodat u op de hoogte blijft. U krijgt dan ook andere vragen en antwoorden te zien.

Abonneren

Abonneer u voor het ontvangen van emails voor uw Gossen Metrawatt METRALINE ISO-CHECK bij:


U ontvangt een email met instructies om u voor één of beide opties in te schrijven.


Ontvang uw handleiding per email

Vul uw emailadres in en ontvang de handleiding van Gossen Metrawatt METRALINE ISO-CHECK in de taal/talen: Engels als bijlage per email.

De handleiding is 0,39 mb groot.

 

U ontvangt de handleiding per email binnen enkele minuten. Als u geen email heeft ontvangen, dan heeft u waarschijnlijk een verkeerd emailadres ingevuld of is uw mailbox te vol. Daarnaast kan het zijn dat uw internetprovider een maximum heeft aan de grootte per email. Omdat hier een handleiding wordt meegestuurd, kan het voorkomen dat de email groter is dan toegestaan bij uw provider.

Andere handleiding(en) van Gossen Metrawatt METRALINE ISO-CHECK

Gossen Metrawatt METRALINE ISO-CHECK Gebruiksaanwijzing - Nederlands - 8 pagina's


Uw handleiding is per email verstuurd. Controleer uw email

Als u niet binnen een kwartier uw email met handleiding ontvangen heeft, kan het zijn dat u een verkeerd emailadres heeft ingevuld of dat uw emailprovider een maximum grootte per email heeft ingesteld die kleiner is dan de grootte van de handleiding.

Er is een email naar u verstuurd om uw inschrijving definitief te maken.

Controleer uw email en volg de aanwijzingen op om uw inschrijving definitief te maken

U heeft geen emailadres opgegeven

Als u de handleiding per email wilt ontvangen, vul dan een geldig emailadres in.

Uw vraag is op deze pagina toegevoegd

Wilt u een email ontvangen bij een antwoord en/of nieuwe vragen? Vul dan hier uw emailadres in.



Info