Suction Unit
www.rossmax.com
IN3V70000000000XX
OBM_IB_SU-V7_
EN_SW_ver2030
Warranty Card
This instrument is covered by a 2 year guarantee from the purchase date. The guarantee
is valid only on presentation of the guarantee card completed by the dealer confirming
purchase date or the receipt. Suction unit accessories are not included. Opening
or altering the instrument invalidates the guarantee. The guarantee does not cover
damage, accidents or noncompliance with the instruction manual. Please contact your
local seller/dealer or www.rossmax.com.
Customer Name: ____________________________________________________
Address: ___________________________________________________________
Telephone: _________________________________________________________
E-mail address: ______________________________________________________
Product Information
Date of purchase: ___________________________________________________
Store where purchased:
____________________________________________________________________
In the event that contaminated uids have moved past the internal unit, remove the tubing
and the lter, clean the casing, and decontaminate the collection canister. Return the device to
your local distributor or authorized Rossmax service center following Rules for returning and
repairing paragraph.
1. Sanitization
Before and after each use, sanitize the collection canister and the connecting tubing.
Rinse each part (9,11,15,15a,16,16a,17,20,22) with clean, hot running water (<50°C) with gentle
dishwashing liquid (non-abrasive). After cleaning, leave the parts to dry in an open, clean
environment.
2. Disinfection
The accessories which can be disinfected are 9,11,15,15a,16,16a,17,20.
Wash with a commercial disinfectant that must be an electrolytic chloroxidizer (active principle:
sodium hypochlorite) specic and follow the instructions and dilution rates supplied by the
disinfectant manufacturer carefully. It is eective on the parts that undergo this treatment only
if the parts to be treated have previously been sanitized. After cleaning, leave the parts to dry in
an open, clean environment.
Parts Material Sanitization Disinfection Steam Sterilization
(Gravity autoclave)
Collection Canister (9) PC(Polycarbonate) O O X
Canister supporting
handle (17) ABS / PC O O X
Canister Lid (15) ABS / PC O O X
Gasket (15a) Silicone O O X
Overflow valve (16) PP O O X
Rubber valve (16a) Silicone O O X
Float (16a) PP O O X
Tubing depository (11) PC O O X
Connecting tubing
(Ø7mm, 180cm) (22) PVC O X X
Connecting tubing
(Ø8mm, 14cm) (20) Silicone O O X
O: Applicable, X: Not applicable
Note:
1. The collection canister and tubing after disinfecting should be dated and stored in clean and dry
areas to avoid recontamination.
2. Do not use any cleaners or disinfectants that contain ammonia, benzene and/or acetone to clean
the parts.
3. Dispose of the solution following the instructions provided by the disinfectant manufacturer.
4. Dispose of the suction catheter in accordance with the local laws and regulations.
Collection canister – 1400 ML
Filter
Silicone and PVC tubing
Maintenance
Rules for returning and repairing
Microbial contamination
Accessories (optional)
Technical Specications
Troubleshooting
Electromagnetic Compatibility Information
Symbols
The collection canister consists of an overow valve, a bottle in PC (polycarbonate) material, a lid
with switch button on handle and must be replaced if there are visible signs of the damage.
Note:
1. Keep the device vertical to allow the overow valve to function correctly.
2. Do not overturn the collection canister during use, in order to prevent the intervention of the
overow valve; should this occur, switch the aspirator o and detach the tubing connected to
the antibacterial lter.
3. To accelerate aspiration and simplify cleaning, it is recommended to put water in the collection
canister before using.
4. Suction at the patient is automatically obstructed when the liquid level reaches the overow
valve located on the underside of the canister lid.
5. Contents of the container should be emptied when automatic overow valve is activated.
Further suctioning could cause damage to the vacuum pump.
The antibacterial lter supplied with Rossmax aspirators must be strictly replaced for each new
patient, or if overow occurs or the lter is saturated. Should the aspirator be used on the same
patient, the lter for single-patient use is recommended to be replaced after every application.
It should also be replaced if the unit is used with a patient whose risk of contamination is unknown.
The lter cannot be sanitized, disinfected, or sterilized.
Note:
1. Do not use the aspirator without the antibacterial lter, the unit could be damaged.
From a bacteriological point of view, it becomes dangerous for the patient.
2. Do not substitute any other material for this antibacterial lter. Substitution may lead to
contamination or poor performance; use only Rossmax lters.
The tubing is recommended to be single use, but the number of cycles of disinfection and/or
cleaning is strictly linked to the employment of the tubing. Therefore, after each cleaning cycle, it
is up to the nal user to verify whether the tubing is suitable for reuse. The component must be
replaced if there are visible signs of the decay of the tubing.
The suction equipment does not need maintenance or lubrication. It is, however to inspect the unit
and accessories before each use.
DANGER:
Electric shock hazard. Do not attempt to open or remove cabinet, there are no user-serviceable
internal components. If service is required, return the device to your local distributor or authorized
Rossmax service center. Opening or tampering with the unit will void warranty.
Rossmax requires the following procedures to be carried out to protect the instrument
and the safety of all who come into contact with it.
Before returning an instrument for repair, the external surfaces and all accessories MUST be
carefully disinfected with a cloth soaked in methylated spirits or hypochlorite-based solution. The
instrument and accessories should then be placed in a bag with a note outlining the disinfection
undertaken. Failure to follow this procedure will result in the instrument being returned to the
purchaser unrepaired. Instruments returned for repair MUST be accompanied by a description of
the problem. Rossmax will not be responsible for damage caused through improper use.
To avoid such damage, please read the instruction carefully.
Rossmax will not be responsible for damage of accessories. Accessories will be replaced at the
customers expense.
In the presence of pathologies with the risk of infection and microbial contamination, we recommend
personal use of the accessories, collection canister, connecting tubing and antibacterial lter and all
accessories should be thoroughly cleaned and disinfected after each use.
Note: Use only with original Rossmax accessories.
Name Model No.
Connecting PVC tubing (Ø7mm-length, 180cm) SU_AC_001_00
Collection canister (1400ml) SU_AC_002_00
Antibacterial lter kit (Antibacterial lter +
2 Silicone tubing Ø8mm-length, 14cm) SU_AC_004_00
Antibacterial lter SU_AC_005_00
Connecting Silicone tubing (Ø8mm-length, 14cm) SU_AC_006_00
Tubing depository SU_AC_008_00
Rechargeable Lithium-ion battery SU_AC_009_00
AC power cord (EU plug) SU_AD_004_00
AC power cord (US plug) SU_AD_005_00
AC power cord (UK plug) SU_AD_006_00
AC power cord (JET plug) SU_AD_007_00
AC/DC switching Adaptor (12V/5A) SU_AD_008_00 (PART No.: UES65-120500SPA1)
Suction class: High Vacuum / High Flow
Adaptor Input: 100-240V, 50/60Hz; Output: DC 12V 5.0A
Max. Vacuum @ pump inlet: -80 kPa/ -600 mmHg (mmHg for reference only)
Max air ow @ pump inlet: 30L/min (normal operation) and 25L/min (power saving/ economical
mode)
Power Consumption: <55W
Collection canister capacity: 1,400 ml
Battery charge duration: approx. 45 mins under normal operation / approx. 60 mins under
power saving/ economical mode at a zero vacuum level (free ow).
Battery charging time: 5 hours (at temp. 25°C)
Battery Rating: 7.2V 2900mAh*2 41.76WH
Weight: Approx. 3.6 (Kgs)
Machine dimension: 363 (W) × 175 (D) × 254.5 (H) mm (Incl. rubber pad)
Operating conditions: Temperature Min. 10°C; Max. 40°C; Relative humidity : 30% ~75%,
non-condensing; Atmospheric pressure: Min. 700 hPa; Max. 1060 hPa
Storage and transportation
conditions: Temperature Min. -20°C; Max. 50°C; Relative humidity: 10~85%, non-
condensing; Atmospheric pressure: Min. 700 hPa; Max. 1060 hPa
IP Classication: IP21: Protection against harmful ingress of water and particulate
matter
If, after checking the described conditions below the device is still not working properly, we
recommend that you consult your local distributor or authorized Rossmax service center.
Problem Possible cause Solution
The device does
not work
The AC adaptor has not been correctly
inserted into the socket of the device
or the power socket
Correctly insert the AC adaptor into
the sockets
Power switch has not been activated. Check if the power switch is turned to
the “I” (ON) position
The unit does
not work from its
battery
Battery at Fully charge the battery
Battery is not installed or inserted
properly Reinstall battery and check if the
battery connector is inserted properly
Battery ages If the unit does not operate after
charging, replace the battery with
new one
Battery broken
If the battery cannot be fully charged
and the GREEN charging indicator
keeps ashing, replace the battery
with new one.
Lack of suction
The collection canister lid has not
been correctly positioned on the
canister
Correctly position the collection
canister lid
Lid gasket (rubber O-ring) not in place Correctly position the gasket on the lid
Leaks or cracks in canister assembly Replace the canister
Filter blocked Replace the filter
Blocked float Incrustation on the float
Remove the canister lid and
overflow valve and extract the
float. Continue by carrying out the
cleaning procedures as described
in the paragraph Sanitization,
Disinfection operations
Vacuum power
poor and/or
non-existent
Vacuum flow adjustment knob/
vacuum regulator completely open Fully close the adjustment knob and
check the vacuum power
Filter blocked Replace the filter
Connecting tubing to the device or
collection canister is clogged, bent,
disconnected or broken; or suction
catheter is twisted or blocked
Check the condition of the tubing,
replace it if blocked or broken and
correctly connect it as per the
Assembling Procedures of this
manual
Canister lid overflow valve closed or
blocked
Unblock the overflow valve, keep the
device upright
Dirty, blocked or damaged pump
Take the device to your local
distributor or authorized Rossmax
service center
1. This device needs to be installed and put into service in accordance with the information provided
in the user manual.
2. WARNING: Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of the V7,
including cables specied by the manufacturer. Otherwise, degradation of the performance of this
device could result.
If higher IMMUNITY TEST LEVELS than those specied in Table 9 are used, the minimum separation
distance may be lowered. Lower minimum separation distances shall be calculated using the
equation specied in 8.10.
Manufacturers declaration-electromagnetic immunity
The V7 is intended for use in the electromagnetic environment specied below. The customer or
the user of the V7 should assure that is used in such and environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment-guidance
Conducted RF
IEC 61000-4-6
3 Vrms:
0,15 MHz – 80 MHz
6 Vrms: in ISM and
amateur radio
bands between
0,15 MHz and 80
MHz
80 % AM at 1 kHz
3 Vrms:
0,15 MHz – 80
MHz
6 Vrms: in ISM and
amateur radio
bands between
0,15 MHz and 80
MHz
80 % AM at 1 kHz
Portable and mobile RF communications
equipment should be used no closer to any
part of the V7 including cables, than the
recommended separation distance calcu-
lated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance:
d = 1,2 √P, d = 1,2 √P 80MHz to 800 MHz, d
= 2,3 √P 800MHz to 2,7 GHz
Where P is the maximum output power rat-
ing of the transmitter in watts (W) accord-
ing to the transmitter manufacturer and d
is the recommended separation distance in
metres (m).
Interference may occur in the vicinity of
equipment marked with the following sym-
bol:
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
NOTE1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2: These guidelines may not apply in all situations. Electromagnetic propagation is aected by
absorption and reection from structures, objects and people.
More information on EMC compliance of the device can be obtained from Rossmax website: www.
rossmax.com.
Class II device
Type BF applied part
Attention: check the instructions before using this device
+More vacuum
SN Device serial number
Risk: Electrocution.
Consequence: Death.
Do not use the device while taking a bath or a shower
ΙSwitch on
ΟSwitch o
Direct current
1639 CE Marking medical ref. Dir 93/42/EEC amended by 2007/47/EC
Date of manufacture
Manufacturer
Attention: Consult accompanying documents
WARNING: The symbol on this product means that its an electronic product and fol-
lowing the European directive 2012/19/EU the electronic products have to be dis-
posed on your local recycling centre for safe treatment.
Rossmax Swiss GmbH,
Widnauerstrasse 1, CH-9435 Heerbrugg, Switzerland
1

Hulp nodig? Stel uw vraag in het forum

Spelregels

Misbruik melden

Gebruikershandleiding.com neemt misbruik van zijn services uitermate serieus. U kunt hieronder aangeven waarom deze vraag ongepast is. Wij controleren de vraag en zonodig wordt deze verwijderd.

Product:

Bijvoorbeeld antisemitische inhoud, racistische inhoud, of materiaal dat gewelddadige fysieke handelingen tot gevolg kan hebben.

Bijvoorbeeld een creditcardnummer, een persoonlijk identificatienummer, of een geheim adres. E-mailadressen en volledige namen worden niet als privégegevens beschouwd.

Spelregels forum

Om tot zinvolle vragen te komen hanteren wij de volgende spelregels:

Belangrijk! Als er een antwoord wordt gegeven op uw vraag, dan is het voor de gever van het antwoord nuttig om te weten als u er wel (of niet) mee geholpen bent! Wij vragen u dus ook te reageren op een antwoord.

Belangrijk! Antwoorden worden ook per e-mail naar abonnees gestuurd. Laat uw emailadres achter op deze site, zodat u op de hoogte blijft. U krijgt dan ook andere vragen en antwoorden te zien.

Abonneren

Abonneer u voor het ontvangen van emails voor uw Rossmax V7 bij:


U ontvangt een email met instructies om u voor één of beide opties in te schrijven.


Ontvang uw handleiding per email

Vul uw emailadres in en ontvang de handleiding van Rossmax V7 in de taal/talen: Engels als bijlage per email.

De handleiding is 1 mb groot.

 

U ontvangt de handleiding per email binnen enkele minuten. Als u geen email heeft ontvangen, dan heeft u waarschijnlijk een verkeerd emailadres ingevuld of is uw mailbox te vol. Daarnaast kan het zijn dat uw internetprovider een maximum heeft aan de grootte per email. Omdat hier een handleiding wordt meegestuurd, kan het voorkomen dat de email groter is dan toegestaan bij uw provider.

Uw handleiding is per email verstuurd. Controleer uw email

Als u niet binnen een kwartier uw email met handleiding ontvangen heeft, kan het zijn dat u een verkeerd emailadres heeft ingevuld of dat uw emailprovider een maximum grootte per email heeft ingesteld die kleiner is dan de grootte van de handleiding.

Er is een email naar u verstuurd om uw inschrijving definitief te maken.

Controleer uw email en volg de aanwijzingen op om uw inschrijving definitief te maken

U heeft geen emailadres opgegeven

Als u de handleiding per email wilt ontvangen, vul dan een geldig emailadres in.

Uw vraag is op deze pagina toegevoegd

Wilt u een email ontvangen bij een antwoord en/of nieuwe vragen? Vul dan hier uw emailadres in.



Info