Introduction
ACT (Artery Check Technology)
Error code for your reference
Installing Batteries
Attaching the lanyard
How to use
Rossmax Fingertip Pulse Oximeter SB200 is used to measure
arterial oxygen saturation (% SpO2) of hemoglobin and pulse
rate, an important indicator of your respiratory function. It is
non-invasive device intended for spot-check of adult and
pediatric whose age is over 3 at home, hospital and clinics.
Attention: Consult the accompanying documents. Please
read this manual carefully before use. Please be sure to
keep this manual.
ACT processes the SpO2 signal and determines the elasticity
of blood vessel based on the derived wave form. It further
classifies the vascular status into 6 levels and presents the result
in an intuitive graphical interface.
Artery and blood circulation in excellent con-
dition.
Artery and blood circulation in good to aver-
age condition.
Artery and blood circulation in average condi-
tion.
Artery and blood circulation in below average
condition.
Artery and blood circulation in poor condition.
Artery and blood circulation in critical condi-
tion.
Note: the classification of artery and blood circulation
condition is for reference only, Please consult with
your physician for further advice.
SENSOR ERROR:
Er
Sensor cannot be detected, return the
device to your local distributor or ser-
vice centre.
MEASURE ERROR:
Er
Signals cannot be detected, turn the de-
vice off and measure again.
AC ERROR:
Er
Weak signal for artery check, turn the
device off and measure again.
1. Use thumb to slide battery cover out.
2. Insert or replace 2 AAA” sized batteries down with the correct
electrical polarity.
You need to replace the batteries when:
1. Low battery icon appears on display
2. The function button is pressed and nothing appears on
display.
Caution: Batteries may leak or explode if used or disposed of
improperly. Remove batteries if the device will be
stored for long time. Do not use different types or
brands of batteries at the same time. Do not mix fully
charged and partially charged batteries at the same
time. The device will automatically shut down in low
battery.
1. Insert the narrow end of the lanyard through the holder.
2. Draw the other end of the lanyard through the loop at the
narrow end and tighten.
1. Open the clip; press the Power On/Off button as .
2. Information of software version appears. Insert one finger(left
hand middle finger is recommended), nail side up, into the
finger opening of the pulse oximeter as .
Note: If no finger insert, the device will auto shut off after 30
seconds.
3. The display shows , pulse oximeter begins its measurement.
Note: Make sure the finger is lying flat. Do not shake and keep
body steady during measurement.
Note: If artery check result cannot be detected, "" will
appear on the screen.
5. Remove the finger, the screen will show artery check level
as .
6. Press button shortly to switch the display to the 3 parameters
(artery check, SpO2 and pulse rate) as .
Note:
1. The SpO2 sensor and photoelectric receiving tube should be
arranged in a way with the subjects arteriole in a position
there between.
2. Make sure the optical path is free from any optical obstacles
like rubberized fabric.
3. Excessive ambient light may affect the measuring result. It
includes fluorescent lamp, dual ruby light, infrared heater,
direct sunlight and etc.
4. Strenuous action of the subject or extreme electrosurgical
interference may also affect the accuracy.
5. The device has a visual and audio signal when the
measurement of SpO2 is lower than 90%.
Cautionary Notes
- This device is to be operated by trained personnel only.
- The device has intended only for spot-checking, but not
medical result evaluation.
- This device is designed to determine the percentage of arterial
oxygen saturation of functional hemoglobin. Factors that may
degrade pulse oximeter performance or affect the accuracy of
the measurement include the following:
Do not apply the pulse oximeter on the same arm as a blood
pressure cuff, arterial catheter or infusion line(s).
Excessive light, such as sunlight or direct home lighting.
Not steady at the site of application (e.g. trembling).
Moisture in the device.
Improperly applied device.
Finger is too large or too small to fit into the device.
Poor pulse quality.
Venous pulsations.
Anemia or low hemoglobin concentrations.
Cardiogreen and other intravascular dyes.
Carboxyhemoglobin.
Methemoglobin.
Dysfunctional hemoglobin.
Artificial nails or fingernail polish.
On fingers with anatomical changes, oedemas, scars or burns.
Name/ Functions of each part
1. OLED display
2. Pulse strength
3. Pulse search icon
4. Artery check icon
5. Alarm icon
1
2
3
4
5
6
7
10
6. Battery indicator
7. Power On/O Button
8. SpO2 icon
9. Pulse icon
10. Battery cover
8
9
4. Your SpO2 and pulse rate values will appear on the screen
after few seconds and artery check result will appear on
screen after 30 - 60 seconds as .
RI_IB_SB200(cont)_EN_2Yr_TP_ver2036.indd 2 2020/9/24 下午 04:13:58
2

Hulp nodig? Stel uw vraag in het forum

Spelregels

Misbruik melden

Gebruikershandleiding.com neemt misbruik van zijn services uitermate serieus. U kunt hieronder aangeven waarom deze vraag ongepast is. Wij controleren de vraag en zonodig wordt deze verwijderd.

Product:

Bijvoorbeeld antisemitische inhoud, racistische inhoud, of materiaal dat gewelddadige fysieke handelingen tot gevolg kan hebben.

Bijvoorbeeld een creditcardnummer, een persoonlijk identificatienummer, of een geheim adres. E-mailadressen en volledige namen worden niet als privégegevens beschouwd.

Spelregels forum

Om tot zinvolle vragen te komen hanteren wij de volgende spelregels:

Belangrijk! Als er een antwoord wordt gegeven op uw vraag, dan is het voor de gever van het antwoord nuttig om te weten als u er wel (of niet) mee geholpen bent! Wij vragen u dus ook te reageren op een antwoord.

Belangrijk! Antwoorden worden ook per e-mail naar abonnees gestuurd. Laat uw emailadres achter op deze site, zodat u op de hoogte blijft. U krijgt dan ook andere vragen en antwoorden te zien.

Abonneren

Abonneer u voor het ontvangen van emails voor uw Rossmax SB200 bij:


U ontvangt een email met instructies om u voor één of beide opties in te schrijven.


Ontvang uw handleiding per email

Vul uw emailadres in en ontvang de handleiding van Rossmax SB200 in de taal/talen: Engels als bijlage per email.

De handleiding is 3.1 mb groot.

 

U ontvangt de handleiding per email binnen enkele minuten. Als u geen email heeft ontvangen, dan heeft u waarschijnlijk een verkeerd emailadres ingevuld of is uw mailbox te vol. Daarnaast kan het zijn dat uw internetprovider een maximum heeft aan de grootte per email. Omdat hier een handleiding wordt meegestuurd, kan het voorkomen dat de email groter is dan toegestaan bij uw provider.

Uw handleiding is per email verstuurd. Controleer uw email

Als u niet binnen een kwartier uw email met handleiding ontvangen heeft, kan het zijn dat u een verkeerd emailadres heeft ingevuld of dat uw emailprovider een maximum grootte per email heeft ingesteld die kleiner is dan de grootte van de handleiding.

Er is een email naar u verstuurd om uw inschrijving definitief te maken.

Controleer uw email en volg de aanwijzingen op om uw inschrijving definitief te maken

U heeft geen emailadres opgegeven

Als u de handleiding per email wilt ontvangen, vul dan een geldig emailadres in.

Uw vraag is op deze pagina toegevoegd

Wilt u een email ontvangen bij een antwoord en/of nieuwe vragen? Vul dan hier uw emailadres in.



Info