741709
1
Verklein
Vergroot
Pagina terug
1/1
Pagina verder
HEM-7113-AP_C_M03_120219.pdf
Instruction Manual
Introduction
Important Safety Information
1. Overview
Model HEM-7113
The unit also stores up to 14 measurements in memory.
Main unit:
A. Display
C. Memory button
D. Battery compartment
E. Air jack
A
B C
D
E
5337886-7A
Thank you for purchasing the OMRON HEM-7113 Digital Automatic Blood
Pressure Monitor.
The OMRON HEM-7113 is a fully automatic blood pressure monitor, operating
on the oscillometric principle. It measures your blood pressure and pulse rate
simply and quickly.
Please read this instruction manual thoroughly before using the unit. For
specific
information about your own blood pressure, CONSULT YOUR
DOCTOR.
Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia or
arteriosclerosis.
Please read this section carefully before using the unit.
Warning:
• Indicates a potentially hazardous situation which, if not avoided, could result in
death or serious injury.
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results and
self-treatment are dangerous.
• People with severe blood flow problems, or blood disorders, should consult a
doctor before using the unit, as cuff inflation can cause internal bleeding.
(Battery Usage)
• If battery fluid should get in your eyes, immediately rinse with plenty of clean
water. Consult a doctor immediately.
Caution:
• Indicates a potentially hazardous situation which, if not avoided, may result in
minor or moderate injury to the user or patient or damage to the equipment or
other property.
(General Usage)
• Do not leave the unit unattended with infants or persons who cannot express
their consent.
• Do not use the unit for any purpose other than measuring blood pressure.
• Do not disassemble the unit or arm cuff.
• Use only the approved arm cuff for this unit. Use of other arm cuffs may result in
incorrect measurement results.
• Make sure that the air tube is not wrapped around other parts of your body when
taking measurements. This could result in injury when the air pressure in the air
tube is increased.
• Do not leave the cuff wrapped on the arm if taking measurements during the
night. This could result in injury.
• Do not inflate the arm cuff over 299 mmHg.
• Do not use a mobile phone or other devices that emit electromagnetic fields,
near the unit. This may result in incorrect operation of the unit.
• Do not operate the unit in a moving vehicle (car, airplane).
• To inflate the cuff manually, refer to Chapter 3.3. If the cuff is over inflated, it can
cause internal bleeding.
(Battery Usage)
• If battery fluid should get on your skin or clothing, immediately rinse with plenty of
clean water.
• Use only four“ AAA” alkaline batteries with this unit. Do not use other types of
batteries.
• Do not insert the batteries with their polarities incorrectly aligned.
• Replace old batteries with new ones immediately. Replace all four batteries at
the same time.
• Remove the batteries if the unit will not be used for three months or more.
• Do not use new and used batteries together. General Precautions
• Do not apply strong shocks and vibrations to or drop the unit and arm cuff.
• Do not take measurements after bathing, drinking alcohol, smoking, exercising
or eating.
• Do not forcibly bend the arm cuff or bend the air tube excessively.
• When removing the air tube, pull on the air plug at the connection with the main
unit not the tube itself.
• Do not inflate the arm cuff when it is not wrapped around your arm.
• Do not wash the arm cuff or immerse it in water.
• Read and follow the“ Important information regarding Electro Magnetic
Compatibility (EMC)” in the Technical Data Section.
• Read and follow the“ Correct Disposal of This Product” in the Technical Data
Section when disposing of the device and any used accessories or optional
parts.
B. START/STOP button.
Automatic
Blood Pressure Monitor
Be sure to wrap the arm cuff correctly so that you get accurate results.
Measurements can be taken in light clothing. However, please remove
thick clothes, such as sweaters, before taking a reading.
2.1 Installing/Replacing the Batteries
2. Preparation
3. Using the Unit
2.2 Battery Life & Replacement
3.1 Applying the Arm Cuff
1. Turn the main unit upside down.
2. Slide the battery cover in the direction
of the arrow while pressing the ribbed
part of the cover.
3. Install or replace four “AAA” size
batteries so that the + (positive)
and - (negative) polarities match
the polarities indicated on the battery
compartment.
4. Put the battery cover back in place.
Slide the battery cover as indicated,
until it clicks into place.
- When the battery low symbol ( ) starts to blink, you will still be able to use
the unit for a short while. You should replace the batteries with new ones
ahead of time.
- When the symbol ( ) remains lit, the batteries are exhausted. You should
replace the batteries with new ones at once. Turn the unit off before replacing
the batteries.
Four new “AAA” alkaline batteries will last for approximately 300 measurements,
when used to take two measurements a day.
Since the supplied batteries are for monitoring use only, they may have
a shorter life and not last for 300 measurements.
• Remove the batteries if the unit will not be used for three months or more.
• Dispose of batteries according to applicable local regulations.
If the battery low symbol ( ) appears on the display,
replace all four batteries at the same time.
1.
1.
To delete stored readings, first press the Memory (M)
button.
Then while holding it down, press the START/STOP
button simultaneously for about 2-3 seconds. All readings
will then be deleted.
2. Press the START/STOP button to turn the unit off.
If you forget to turn the unit off, it will automatically shut
itself off after five minutes.
To Delete All the Values Stored in Memory
You cannot delete individual stored readings, all the
readings in the unit will be deleted.
2
1
4. Handling Errors and Problems
4.1 Error Messages
Error Display Cause Remedy
Cuff is under inflated.
Carefully read and repeat the
steps listed in section 3.3.
Movement during
measurement.
Repeat measurement. Remain
still and do not talk during
measurement.
Air plug disconnected.
Insert the air plug securely.
Refer to section 3.1.
Arm cuff not applied
correctly.
Apply the arm cuff correctly.
Refer to section 3.1.
Clothing is interfering
with the arm cuff.
Remove any clothing
interfering with the arm cuff.
Air is leaking from the
arm cuff.
Replace cuff with new one.
Arm cuff has been
inflated above
299 mmHg.
Do not inflate the arm cuff
above 299 mmHg.
Refer to section 3.3.
Intended Use
This product is designed to measure the blood pressure and pulse rate of
people within the range of the designated arm cuff, following the instructions in
this instruction manual.
It is mainly designed for general household use. Please read the Important
Safety Information in this instruction manual before using the unit.
1 - 2 cm
2.
Note:
You can take a measurement on either your left or right arm. The blood
pressure can differ between the right arm and the left arm and therefore
also the measured blood pressure values can be different. OMRON
recommends to always use the same arm for measurement. If the
values between the two arms differ substantially, please check with your
physician which arm to use for your measurement.
Insert the air plug into the air jack
on the left side of the main unit.
Apply the arm cuff to your upper arm. The air tube should be
centred on the inside of your arm and point down the inside of the
arm, so that the air tube runs down the inside of your forearm and
is in line with your middle finger.
3.3 Instructions for Special Conditions
5. Press the START/STOP button to turn the unit off.
1. Press the START/STOP button
to turn the unit on.
Measurement starts.
Note: If you forget to turn the unit off, it will automatically shut itself off after
five minutes.
If your systolic pressure is known to be more than 220 mmHg, press
and hold the START/STOP button until the arm cuff inflates 30 to
40 mmHg higher than your suspected systolic pressure.
Arm cuff:
F. Arm cuff
(Medium cuff:
arm circumference 22-32 cm)
G. Air tube
Note:
GF
H
Package contents:
Display:
J. Systolic blood pressure
K. Diastolic blood pressure
L. Memory symbol
N. Heartbeat symbol
1. Flashes during measurement
O. Pulse display
P. Deflation symbol
Q. Battery low symbol
I.
Instruction manual
Blood pressure monitor unit
I
J
L
N
P
K
O
Q
H. Air plug
M. Irregular Heartbeat symbol
M
If air starts to leak from the arm cuff,
please replace with a new one.
Refer to “6. Optional Parts”.
The optional arm cuff does not
include the air plug. Do not throw
the air plug away in case
purchasing the new one.
Four “AAA” alkaline (LR03)
batteries
2. If flashing after measurement
complete indicates blood pressure
out of recommended range
Note: If your systolic or diastolic
pressure is outside the
standard range
(above 135/85 mmHg),
the Heartbeat symbol ( )
will blink. Refer to
“3.2 Taking a Reading”.
Important:
4. Remove the arm cuff.
Your blood pressure monitor includes an
irregular heartbeat feature. Irregular heart-
beats can influence the results of the
measurement. The irregular heartbeat
algorithm automatically determines if the
measurement is usable or needs to be
repeated. If the measurement results are affected by
irregular heartbeats but the result is valid, the result is
shown together with the irregular heartbeat symbol ( ).
If the irregular heartbeats cause the measurement to be
invalid, no result is shown. If the Irregular heartbeat
symbol ( ) is shown after you have taken a measure-
ment, repeat the measurement. If the Irregular heartbeat
symbol ( ) is shown frequently, please make your doctor
aware of it.
NormalHeartbeat
Blood
pressure
Blood
pressure
Pulse
Pulse
Warning:
What is Arrhythmia?
An irregular heartbeat is a heartbeat
rhythm that varies by more than 25%
from the average heartbeat rhythm
detected while the unit is measuring
the systolic and diastolic blood
pressure. If such an irregular rhythm
is detected more than twice during
measurement, the irregular heartbeat
symbol ( ) appears on the symbol
when the measurement results are
displayed.
A heartbeat is stimulated by electrical signals that cause the heart to contract.
Arrhythmia is a condition where the heartbeat rhythm is abnormal due to flaws
in the bio-electrical system that drives the heartbeat. Typical symptoms are
skipped heartbeats, premature contraction, an abnormally rapid (tachycardia)
or slow (bradycardia) pulse. This can be caused by heart disease, aging,
physical predisposition, stress, lack of sleep, fatigue etc. Arrhythmia can only
be diagnosed by a doctor through a special examination. Whether the
appearance of the irregular heartbeat symbol ( ) in the results indicates
arrhythmia or not can only be determined by an examination and diagnosis by
your doctor.
What is Irregular Heartbeat?
If the irregular heartbeat symbol ( ) is shown frequently, please make
your doctor aware of it. Conducting self-diagnosis and treatment based on
measurement results is dangerous. Be sure to follow the instructions of
your doctor.
Irregular
Short
Heartbeat
Long
Note: The measurement values
continue to be stored in
memory even after the
batteries are replaced.
Arm cuff pressure
does not rise.
Make sure that the air
jack is connected
securely.
Refer to section 3.1.
Air is leaking from the
arm cuff.
Replace the arm cuff
with a new one.
Arm cuff deflates too
soon.
The arm cuff is loose.
Apply the cuff correctly
so that it is firmly
wrapped around the
arm.
Refer to section 3.1.
Cannot measure or
readings are too low
or too high.
Inflate the cuff so that it
is 30 to 40 mmHg
above your previous
measurement result.
Refer to section 3.3.
The unit loses power
during measurement.
The batteries are
empty.
Replace the batteries
with new ones.
Nothing happens
when you press the
buttons.
The batteries are
empty.
Replace the batteries
with new ones.
The batteries have
been inserted
incorrectly.
Insert the batteries with
the correct (+/ -)
polarity.
4.2 Troubleshooting
Problem Cause Remedy
The reading is
extremely low (or
high).
Arm cuff not applied
correctly.
Apply the arm cuff
correctly.
Refer to section 3.1.
Movement or talking
during measurement.
Remain still and do not
talk during
measurement.
Clothing is interfering
with the arm cuff.
Remove any clothing
interfering with the arm
cuff.
Other problems.
Press the START/STOP button and repeat
measurement.
If the problem continues, try replacing the
batteries with new ones.
If this still does not solve the problem, contact your
local authorized OMRON distributor or dealer.
Note: If the memory is full, the monitor will delete the oldest readings.
This unit has a memory capable of storing 14 sets of readings.
Press the Memory (M) button.
The result from the most recent measurement is displayed.
Note: If there are no measurements results
stored in memory, the screen to the right
is displayed.
If your systolic or diastolic pressure is
outside the standard range, the heartbeat
symbol will blink when the measurement
result is displayed.
Refer to “3.2 Taking a Reading”.
Press the Memory (M) button repeatedly to
cycle through the previous measurement
results.
The Memory number appears for a second
before the pulse rate is displayed.
The newest set is numbered “1”.
Press the START/STOP button to turn the unit off.
If you forget to turn the unit off, it will automatically shut
itself off after five minutes.
3.4 Using the Memory Function
Important:
1.
2.
3.
At the end of the measurement, if an irregular heartbeat
is detected, the irregular heartbeat symbol will be
displayed with the result.
1 to 2 cm
3.2 Taking a Reading
Taking a Reading on the Right Arm
3.
Sit on a chair with your feet flat on
the floor and place your arm on a
table so that the arm cuff will be
at the same level as your heart.
Keep still and do not talk during measurement.
2.
Press the START/STOP button to turn on the unit.
1.
3.
Check the measurement results.
The unit automatically stores blood
pressure and pulse rate into its memory.
Refer to “3.4 Using the Memory
Function”.
Notes:
• Self-diagnosis of measured results and treatment are dangerous.
Please follow the instructions of your doctor.
• Wait 2-3 minutes before taking another blood pressure reading.
Waiting between readings allows the arteries to return to the condition prior
to taking the blood pressure measurement.
Diastolic blood
pressure
Systolic blood
pressure
Pulse display
When the cuff is positioned correctly, close the fabric fastener
firmly.
Apply the cuff so that the air tube is
at the side of your elbow.
Notes:
• Be careful not to rest your arm on
the air tube, or otherwise restrict the
flow of air to the cuff.
• The cuff should be 1 to 2 cm above
the elbow.
Note: To cancel a measurement, press the START/STOP button to turn off
the unit and to release the air in the arm cuff.
Error Display Cause Remedy
Battery power is low.
Replace all four “AAA”
batteries with new ones.
Refer to section 2.1.
Device error.
Blinks or appears
continuously
Note: The irregular heartbeat symbol ( ) may also be displayed with error messages.
Note: You cannot inflate the cuff above 299 mmHg. (An error will be displayed
if you try to inflate the cuff above 299 mmHg.)
Note: Do not apply more pressure than necessary.
2. When the cuff starts to inflate, press the START/STOP button
and keep it pressed until the pressure is 30 to 40 mmHg
higher than your suspected systolic pressure.
3. Release the START/STOP button when the cuff has been
inflated to the desired pressure.
The cuff starts to deflate and measurement starts.
4. The rest of the procedure is the same as for normal
measurement. Refer to “3.2 Taking a Reading”, steps 3 to 5.
The air plug is not
securely inserted into
the main unit.
The arm cuff has
not been inflated
sufficiently.
5. Maintenance and Storage
Maintenance
To protect your unit from damage, please observe the following:
7. Technical Data
6. Optional Parts
• The unit should be cleaned with a soft, dry cloth.
Use a soft, moistened cloth and soap to clean the arm cuff.
Calibration and Service
• The accuracy of this blood pressure monitor has been
carefully tested and is designed for a long service life.
Note: Do not bend the air tube excessively.
Unplug the air tube from the air jack.
Gently fold the air tube
into the arm cuff.
Storage
Do not store the unit in the following situations:
• If the unit is wet.
• Locations exposed to extreme temperatures, humidity, direct sunlight,
dust or corrosive vapours.
• Locations exposed to vibrations, shocks or where it will be at an angle.
Medium arm cuff
Arm circumference
22 - 32 cm
Large arm cuff
Arm circumference
32 - 42 cm
• Do not subject the main unit and cuff to extreme
temperatures, humidity, moisture or direct sunlight.
• Do not fold the cuff or tubing tightly.
• Do not disassemble the unit.
• Do not subject the unit to strong shocks or vibrations
(for example, dropping the unit on the floor).
• Do not use volatile liquids to clean the main unit.
• Do not wash the arm cuff or immerse it in water.
• Do not use petrol, thinners or similar solvents to clean
the arm cuff.
• Do not carry out repairs of any kind by yourself.
If a defect occurs, consult your local authorized
OMRON distributor or dealer.
It is generally recommended to have the unit inspected every
two years to ensure correct functioning and accuracy. Please
consult your local authorized OMRON distributor or dealer.
Automatic Blood Pressure Monitor
HEM-7113 (HEM-7113-AP)
LCD Digital Display
Oscillometric method
Pressure: 0 mmHg to 299 mmHg
Pulse: 40 to 180/min.
14 Measurements
Pressure: ±3 mmHg
Pulse: ±5% of display reading
Automatic by electric pump
Automatic pressure release valve
4 "AAA" batteries 1.5V
Capacity of new alkaline batteries is approximately
300 measurements
Notes:
• Subject to technical modification without prior notice.
Disposal of this product and used batteries should be carried out in accordance
with the national regulations for the disposal of electronic products.
Operating
temperature/ Humidity
Storage temperature/
Humidity/ Air Pressure
Console Weight
Cuff Weight
Outer Dimensions
Cuff Dimensions
Package Content
10°C to 40°C/ 30 to 85% RH
-20°C to 60°C/ 10 to 95% RH/ 700-1060 hPa
Approximately 245 g without batteries
Approximately 120 g
Approximately
104 (w) mm × 64 (h) mm × 128.5 (I) mm
Approximately 146 mm × 446 mm
(Medium cuff: arm circumference 22 to 32 cm)
Main unit, cuff, Instruction manual, Battery set
Product Description
Model
Display
Measurement Method
Measurement Range
Memory
Accuracy
Inflation
Deflation
Power Source
Battery life
Applied Part
Protection Against
Electric Shock
= Type B
Internally powered ME equipment
Asia Pacific HQ
Production Facility
OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto,
617-0002 JAPAN
OMRON HEALTHCARE SINGAPORE PTE LTD.
OMRON HEALTHCARE MANUFACTURING
VIETNAM CO., LTD.
Binh Duong Province, VIETNAM
438A Alexandra Road, #05-05/08, Alexandra Technopark,
Singapore 119967
www.omron-healthcare.com.sg
This OMRON product is manufactured under the strict quality system of OMRON
HEALTHCARE Co., Ltd., Japan. The Core component for OMRON blood pressure
monitors, which is the Pressure Sensor, is produced in Japan for assembly.
Contact your local
authorized OMRON
distributor or dealer

Hulp nodig? Stel uw vraag in het forum

Spelregels

Misbruik melden

Gebruikershandleiding.com neemt misbruik van zijn services uitermate serieus. U kunt hieronder aangeven waarom deze vraag ongepast is. Wij controleren de vraag en zonodig wordt deze verwijderd.

Product:

Bijvoorbeeld antisemitische inhoud, racistische inhoud, of materiaal dat gewelddadige fysieke handelingen tot gevolg kan hebben.

Bijvoorbeeld een creditcardnummer, een persoonlijk identificatienummer, of een geheim adres. E-mailadressen en volledige namen worden niet als privégegevens beschouwd.

Spelregels forum

Om tot zinvolle vragen te komen hanteren wij de volgende spelregels:

Belangrijk! Als er een antwoord wordt gegeven op uw vraag, dan is het voor de gever van het antwoord nuttig om te weten als u er wel (of niet) mee geholpen bent! Wij vragen u dus ook te reageren op een antwoord.

Belangrijk! Antwoorden worden ook per e-mail naar abonnees gestuurd. Laat uw emailadres achter op deze site, zodat u op de hoogte blijft. U krijgt dan ook andere vragen en antwoorden te zien.

Abonneren

Abonneer u voor het ontvangen van emails voor uw Omron HEM-7113 bij:


U ontvangt een email met instructies om u voor één of beide opties in te schrijven.


Ontvang uw handleiding per email

Vul uw emailadres in en ontvang de handleiding van Omron HEM-7113 in de taal/talen: Engels als bijlage per email.

De handleiding is 1,83 mb groot.

 

U ontvangt de handleiding per email binnen enkele minuten. Als u geen email heeft ontvangen, dan heeft u waarschijnlijk een verkeerd emailadres ingevuld of is uw mailbox te vol. Daarnaast kan het zijn dat uw internetprovider een maximum heeft aan de grootte per email. Omdat hier een handleiding wordt meegestuurd, kan het voorkomen dat de email groter is dan toegestaan bij uw provider.

Uw handleiding is per email verstuurd. Controleer uw email

Als u niet binnen een kwartier uw email met handleiding ontvangen heeft, kan het zijn dat u een verkeerd emailadres heeft ingevuld of dat uw emailprovider een maximum grootte per email heeft ingesteld die kleiner is dan de grootte van de handleiding.

Er is een email naar u verstuurd om uw inschrijving definitief te maken.

Controleer uw email en volg de aanwijzingen op om uw inschrijving definitief te maken

U heeft geen emailadres opgegeven

Als u de handleiding per email wilt ontvangen, vul dan een geldig emailadres in.

Uw vraag is op deze pagina toegevoegd

Wilt u een email ontvangen bij een antwoord en/of nieuwe vragen? Vul dan hier uw emailadres in.



Info