Inserting and changing batteries
1. Push out the tray at the bottom of the blood pressure monitor so that the battery
compartment door is exposed.
2. Press slightly on the handle of door while you lift and remove it. Insert or change the batteries
in the battery compartment. Make sure that the batteries are placed correctly. The door is put
back in its place.
3. The batteries must be changed when the battery symbol on the display shows low battery.
The battery symbol will disappear once the device is switched on and the battery is charged.
As the supplied batteries are for test only, they may be discharged earlier than batteries you buy
in stores. Replace the batteries in pairs. Remove batteries when unit is not in use for extended
periods of time.
Error search
If irregularities occur during use, please check the following items before you contact your vendor:
If the device still does not function, please contact the store where the device was purchased.
Do not attempt to open the device or adjust any of the inner parts yourself.
Establishing arrhythmia (heart rhythm disturbances)
An unstable heart rhythm can lead to or indicate cardiovascular diseases. By means of
advanced technology, this blood pressure device can warn the user that he or she has an
irregular heart rhythm which in turn contributes to better health monitoring.
If the symbol
IHB
appears during blood pressure measurement, it means that an irregular heart
rhythm has been established. This should only be regarded as a warning. Note that it is
important that the user is relaxed, sits still, and does not speak during measurement. It is
strongly recommended that you contact your doctor if the irregular heart rhythm symbol appears
several times in succession (5 times or more).
Memory
This blood pressure device can be used by two users. Each user can store up to 60 measurements.
1. Press .
2. Choose user 1 or 2 by pressing .
3. Press to enter memory.
4. The blood pressure device will first show the average of the three most recent measurements
stored in the memory.
5. Every time is pressed, previously stored measurements are shown. The latest
measurement is shown first.
6. To clear memory press to choose user 1 or 2.
7. Press and hold the button for approximately 5 seconds and all data for the chosen user
are erased automatically.
Recalibration service
All blood pressure devices must be recalibrated to ensure continual measurement accuracy.
The blood pressure device should be recalibrated after 2 years of normal use (approx. 3
measurements a day). Subsequently, the device will show CA on the display. The device should
also be recalibrated if e.g. it has been dropped on the floor or in fluids, if fluid has been spilled over
or on the device, or if it has been exposed to extremely high or low temperatures.
This monitor must only be calibrated by authorised personel and should therefore be sent to
HoMedics Group Ltd for re-calibration. Do not carry out any repairs yourself. Please contact the
shop where you bought the product for further details on the procedure for submitting the monitor or
contact HoMedics customer service on +44 (0) 1732 378557 or email: cservice@homedics.co.uk.
10
GB
GB
Symptom Possible reasons Correction
The device does not switch
on when the stop/start button
is activated.
A change of batteries is required. Change to four new batteries.
EE measurement error
symbol in the display, or
the blood pressure
measurement is excessively
low (or high).
The cuff is fitted incorrectly.
Fit the cuff again so that it is
positioned correctly.
Did you talk or move during
measurement?
Restart measurement and keep
the arm steady during the entire
measurement.
Shaking of the arm while wearing
the cuff.
E1 error symbol on the
display.
Airflow is irregular. The tube is not
properly applied to the device.
Check the cuff’s connection. Restart
measurement.
E2 error symbol on the
display.
The inflation pressure exceeds 300mmHg. Switch off the device and restart
measurement.
E3 error symbol on the
display.
Data error. Remove and reinsert batteries.
CA on the display. Calibration warning. See page 48 for more information.
6

Hulp nodig? Stel uw vraag in het forum

Spelregels

Misbruik melden

Gebruikershandleiding.com neemt misbruik van zijn services uitermate serieus. U kunt hieronder aangeven waarom deze vraag ongepast is. Wij controleren de vraag en zonodig wordt deze verwijderd.

Product:

Bijvoorbeeld antisemitische inhoud, racistische inhoud, of materiaal dat gewelddadige fysieke handelingen tot gevolg kan hebben.

Bijvoorbeeld een creditcardnummer, een persoonlijk identificatienummer, of een geheim adres. E-mailadressen en volledige namen worden niet als privégegevens beschouwd.

Spelregels forum

Om tot zinvolle vragen te komen hanteren wij de volgende spelregels:

Belangrijk! Als er een antwoord wordt gegeven op uw vraag, dan is het voor de gever van het antwoord nuttig om te weten als u er wel (of niet) mee geholpen bent! Wij vragen u dus ook te reageren op een antwoord.

Belangrijk! Antwoorden worden ook per e-mail naar abonnees gestuurd. Laat uw emailadres achter op deze site, zodat u op de hoogte blijft. U krijgt dan ook andere vragen en antwoorden te zien.

Abonneren

Abonneer u voor het ontvangen van emails voor uw Homedics BPA-300 bij:


U ontvangt een email met instructies om u voor één of beide opties in te schrijven.


Ontvang uw handleiding per email

Vul uw emailadres in en ontvang de handleiding van Homedics BPA-300 in de taal/talen: Engels als bijlage per email.

De handleiding is 3,44 mb groot.

 

U ontvangt de handleiding per email binnen enkele minuten. Als u geen email heeft ontvangen, dan heeft u waarschijnlijk een verkeerd emailadres ingevuld of is uw mailbox te vol. Daarnaast kan het zijn dat uw internetprovider een maximum heeft aan de grootte per email. Omdat hier een handleiding wordt meegestuurd, kan het voorkomen dat de email groter is dan toegestaan bij uw provider.

Uw handleiding is per email verstuurd. Controleer uw email

Als u niet binnen een kwartier uw email met handleiding ontvangen heeft, kan het zijn dat u een verkeerd emailadres heeft ingevuld of dat uw emailprovider een maximum grootte per email heeft ingesteld die kleiner is dan de grootte van de handleiding.

Er is een email naar u verstuurd om uw inschrijving definitief te maken.

Controleer uw email en volg de aanwijzingen op om uw inschrijving definitief te maken

U heeft geen emailadres opgegeven

Als u de handleiding per email wilt ontvangen, vul dan een geldig emailadres in.

Uw vraag is op deze pagina toegevoegd

Wilt u een email ontvangen bij een antwoord en/of nieuwe vragen? Vul dan hier uw emailadres in.



Info