ACCESSORIES SUPPLIED
DESCRIPTION
COMPLETE ASPIRATION JAR 1000cc
CONICAL FITTING
TUBES SET 6 mm x 10 mm
ASPIRATION PROBE CH20
ANTIBACTERIAL FILTER
Antibacterial Filter: The filter is produced with (PTFE) hydrophobic material witch prevents fluids entering the pneumatic circuit.
The filter is for a single patient use which will protect patients and machines from cross contamination.
When the filter is wet, it’s not possible to use the unit therefore the filter should be changed immediately.
In case of possible contamination or discolouration, change the filter immediately.
Don’t use the suction unit without the protection filter fitted.
Suction catheter: Single-use device to be used on a single patient. Do not wash or re-sterilize after use. Reuse may cause cross-
infections.
Don’t use after lapse of the sell-by date
WARNING: Suction tubes for insertion in the human body purchased separately from the machine should comply with ISO
10993-1 standards on material biocompatibility.
Aspiration jar: The mechanical resistance of the component is guaranteed up to 30 cycles of cleaning and sterilization. Beyond this
limit, the physical-chemical characteristics of the plastic material may show signs of decay. Therefore, we recommend that you to
change it.
Silicone tubes: the number of cycles of sterilization and/or cleaning is strictly linked to the employment of the said tube. Therefore, after
each cleaning cycle, it is up to the final user to verify whether the tube is suitable for reuse. The component must be replaced if there
are visible signs of decay of the material constituting the said component.
Conical fitting: the number of cycles of sterilization and the number of cleaning cycles is strictly linked to the employment of the said
component. Therefore, after each cleaning cycle, it is up to the final user to verify whether the fitting is suitable for reuse.
The component must be replaced if there are visible signs of decay in the material constituting the said component.
Service life of the device: More than 850 hours of operation (or 3 years) in accordance with the standard conditions of testing and
operation. Shelf life: maximum 5 years from the date of manufacture.
CLEANING OF ACCESSORIES
To clean the plastic housing of the device wear disposable latex gloves and clean with denaturated alcohol or hypoclorite solutions.
Washing and / or cleaning the autoclavable jar as to be carried out as follows:
Wear protection gloves and apron (glasses and face mask if necessary) to avoid contact with contaminating substances;
Disconnect the jar from the device
Disconnect all tubes from the jar and the protection filter
Empty and dispose of the content and of the suction catheter according to the laws in force in your country;
Separate all parts of the cover (overflow valve, o-ring);
After disposing of disposable parts and disassembling the jar wash in running cold water and rinse thoroughly.
Then soak in warm water (temperature shall not exceed 60°C).
Wash thoroughly and if necessary use a non-abrasive brush to remove incrustations.
Rinse in running warm water and dry all parts with a soft cloth (non-abrasive).
The jar and the cover can be autoclaved by placing the parts into the autoclave and running one sterilization stem cycle at 121°C
(1 bar relative pressure) making sure that the jar is positioned upside down.
Mechanical resistance of the jar is guaranteed up to 30 cycles of sterilization and cleaning at the indicated conditions
(EN ISO 10079-1). Beyond this limit the physical-mechanical characteristics of the plastic may decrease and replacement of the part is
therefore recommended.
After sterilization and cooling at environment temperature of the parts make sure that these are not damaged.
Assemble the jar as follows:
Place the overflow valve into its seat in the cover (under VACUUM connector)
Insert floating valve keeping the o-ring towards the opening of the cage
Place the o-ring into its seat around the cover
After completing assembling operations always make sue that cover seals perfectly to avoid vacuum leakages or liquid exit
The aspiration tubes can be sterilized on autoclave using a sterilization cycle at 120°C.
The conical connector can be sterilized on autoclave using a sterilization cycle at 121°C.
The device is ready for a new employment now.
DO NOT WASH, STERILIZE OR PUT IN AUTOCLAVE THE ANTIBACTERIAL FILTER
5

Hulp nodig? Stel uw vraag in het forum

Spelregels
1

Forum

Cami-New-Askir-30

Zoeken resetten

Misbruik melden

Gebruikershandleiding.com neemt misbruik van zijn services uitermate serieus. U kunt hieronder aangeven waarom deze vraag ongepast is. Wij controleren de vraag en zonodig wordt deze verwijderd.

Product:

Bijvoorbeeld antisemitische inhoud, racistische inhoud, of materiaal dat gewelddadige fysieke handelingen tot gevolg kan hebben.

Bijvoorbeeld een creditcardnummer, een persoonlijk identificatienummer, of een geheim adres. E-mailadressen en volledige namen worden niet als privégegevens beschouwd.

Spelregels forum

Om tot zinvolle vragen te komen hanteren wij de volgende spelregels:

Belangrijk! Als er een antwoord wordt gegeven op uw vraag, dan is het voor de gever van het antwoord nuttig om te weten als u er wel (of niet) mee geholpen bent! Wij vragen u dus ook te reageren op een antwoord.

Belangrijk! Antwoorden worden ook per e-mail naar abonnees gestuurd. Laat uw emailadres achter op deze site, zodat u op de hoogte blijft. U krijgt dan ook andere vragen en antwoorden te zien.

Abonneren

Abonneer u voor het ontvangen van emails voor uw Cami New Askir 30 bij:


U ontvangt een email met instructies om u voor één of beide opties in te schrijven.


Ontvang uw handleiding per email

Vul uw emailadres in en ontvang de handleiding van Cami New Askir 30 in de taal/talen: Engels als bijlage per email.

De handleiding is 0,31 mb groot.

 

U ontvangt de handleiding per email binnen enkele minuten. Als u geen email heeft ontvangen, dan heeft u waarschijnlijk een verkeerd emailadres ingevuld of is uw mailbox te vol. Daarnaast kan het zijn dat uw internetprovider een maximum heeft aan de grootte per email. Omdat hier een handleiding wordt meegestuurd, kan het voorkomen dat de email groter is dan toegestaan bij uw provider.

Uw handleiding is per email verstuurd. Controleer uw email

Als u niet binnen een kwartier uw email met handleiding ontvangen heeft, kan het zijn dat u een verkeerd emailadres heeft ingevuld of dat uw emailprovider een maximum grootte per email heeft ingesteld die kleiner is dan de grootte van de handleiding.

Er is een email naar u verstuurd om uw inschrijving definitief te maken.

Controleer uw email en volg de aanwijzingen op om uw inschrijving definitief te maken

U heeft geen emailadres opgegeven

Als u de handleiding per email wilt ontvangen, vul dan een geldig emailadres in.

Uw vraag is op deze pagina toegevoegd

Wilt u een email ontvangen bij een antwoord en/of nieuwe vragen? Vul dan hier uw emailadres in.



Info